Physio-Control All models error codes
Codes
- 5Error codeIP cable ports - P1 and P2 connection ports for invasive pressure cables, which invasively measure arterial blood pressures, central venous pressure (CVP), or intracranial pressure. Note: If device is configured for temperature monitoring, P1 and P2 are replaced by a single port labeled TEMP
- 6Error codeTherapy Processor Error: a. Verify that the appropriate connecting cables and wire harnesses are functional. b. Possible A04 Therapy PCB failure
- 7Error codePower Processor Error: a. Verify that the appropriate connecting cables and wire harnesses are functional. b. Possible A03 Power PCB failure
- 8Error codeCODE button - Prints a summary of the current patient documentation, including vital signs and waveforms. SUMMARY
- 9Error codeNIBP button (optional) - Initiates blood pressure measurement. LED is illuminated when BP measurement is being obtained
- 9xError codePPxx
- 500dError codePC ERROR CONFIG VERSION (System configuration version mismatch; param = value read.) _ _ _
- 500eError codePC ERROR CONFIG CRC (NVRAM system configuration CRC bad; param = value read.) _ _ _
- 501aError codePC ERROR RTC DRIFT (Processor and RTC time out of sync; param = drift.) _ _ _
- 600cError codeECG ERROR DSP VOLTAGE (DSP preamp supply voltage out of range.) _ _ _
- 600dError codeECG ERROR PREAMP CALIBRATION (NVRAM calibration constants out of range. HW unit reported calibration error.) _ _ _
- 600eError codeECG ERROR NVRAM FAULT (NVRAM redundant value mismatch detected. NVRAM possibly bad.) _ _ _
- 900eError codePPSP ERROR CONFIG (SpO2 module detected but not in manufacturing configuration; param: 1 = found Masimo but _ _ should not, 2 = unit not found.)
- 902dError codePPSP ERROR MNC CONFIG B (SpO2 module detected but manufacturing configuration, not configured for Nellcor _ _ _ _ sensor support; param: 1 = Masimo doesn’t support config B, 2 = Device doesn’t support config B.)
- 4011Error codeSM ERROR DEFIB SERVICE SYNC (Failed to synchronize the defib charge after cold boot; param = time since last _ _ _ _ boot.)
- 4012Error codeSM ERROR FONT VOICE CKSUM (Invalid checksum in font/voice. Font/voice checksum error found after cold boot.) _ _ _ _
- 4013Error codeSM ERROR FONT VOICE CRC (Invalid CRC in font/voice flash. Font/voice CRC error; status code = top 16 bits are _ _ _ _ stored CRC, low 16 bits are computed CRC.)
- 5002Error codePC ERROR WATCHDOG SHORT FAILURE (Main watchdog short test failure. Watchdog failure; param: 0=None, _ _ _ _ 1=short, 2=long, 3=power.)
- 5003Error codePC ERROR WATCHDOG LONG FAILURE (Main watchdog long test failure. Watchdog failure; param: 0=None, 1=short, _ _ _ _ 2=long, 3=power.)
- 5004Error codePC ERROR WATCHDOG UNEXPECTED (Unexpected main watchdog reset. Top 16 bits = seconds since last set. Low _ _ _ 16 bits = watchdog status.)
- 5005Error codePC ERROR RAM AT BOOT (RAM error detected during boot; param = Ram Addr of error.) _ _ _ _
- 5006Error codePC ERROR BAD CHECKSUM (Program contents failed Checksum test.) _ _ _
- 5011Error codePC ERROR METERS VERSION (System meters version mismatch; param = value read.) _ _ _
- 5012Error codePC ERROR METERS CRC (NVRAM system meters/counters CRC bad; param = value read.) _ _ _
- 5013Error codePC ERROR MFG DATA VERSION (Manufacturing data version mismatch.) _ _ _ _
- 5014Error codePC ERROR MFG DATA CRC (NVRAM manufacturing data CRC bad.) _ _ _ _
- 5015Error codePC ERROR FORCED RESET FAILED (Forced watchdog reset failed. Unit failed to reset.) _ _ _ _
- 5019Error codePC ERROR RTC BAD (RTC is not running.) _ _ _
- 5032Error codePC ERROR RTC NO BATTERY (No Coin cell battery for the RTC.) _ _ _ _
- 5033Error codePC ERROR RTC READ (RTC read is not consistent.) _ _ _
- 9119Error codePPNI ERROR CONFIG MISMATCH (Device with NIBP module not configured for it; param = SW version.) _ _ _
- 9205Error codePPCO ERROR WRITING TO MODULE (CO2 module detected but not in manufacturing configuration. Unit disabled. _ _ _ _ Param = config-info.)
- 9304Error codePPTI CONFLICT IP TEMPER ERROR CONFIG (IP and Temperature manufacturing configuration bits both on at the _ _ _ _ _ same time.)
- a00aError codeTH ERROR DE WRONG ENG SELECT (Incorrect energy selected; param = energy index.) _ _ _ _ _
- a00bError codeTH ERROR SHOCK NOT DELIVERED (Shock not delivered.) _ _ _ _
- a00dError codeTH ERROR CHARGING EXPIRED (Charging time expired.) _ _ _
- a00fError codeTH ERROR CAP OUT OF RANGE (Capacitor is out of range. Calibration failure.) _ _ _ _ _
- A01Error codeSystem PCB
- a01aError codeTH ERROR PACER FAULT (Pacing fault condition occurred [rate(0), current(1), pulse width(2)], limit exceeded; param = pacer-fault _ _ _ type.)
- a01bError codeTH ERROR DEFIB WDT DISABLE FAIL (Unable to turn off defib WDT; param = ASIC defib ctrl register.) _ _ _ _ _
- a002Error codeTH ERROR DEFIB LINK DOWN (Lost contact with defib processor. Serial communications link between the main and defib processor is _ _ _ _ not functioning.This unit can no longer administer defib therapy. Cycling power may clear the error temporarily, but unit is questionable; param = last defib message time stamp.)
- A02Error codeMemory PCB
- a003Error codeTH ERROR PACER LINK DOWN (Lost contact with pacer processor. Serial communications link between the main and pacer processor is _ _ _ _ not functioning.)
- a004Error codeTH ERROR UNEXPECTED ENERGY (Unexpected energy in the capacitor. No charge delivered.) _ _ _
- a005Error codeTH ERROR CAP OVERCHARGED (Over-charged capacitor.) _ _ _
- a008Error codeTH ERROR DEFIB DISABLE (No communication with defib HW.) _ _ _
- a010Error codeTH ERROR PA RATE OUT OF RANGE (Pacing rate out of tolerance; param: high-16 = pacer selected range, low-16 = pacer actual rate.) _ _ _ _ _ _
- a011Error codeTH ERROR PA CURRENT OUT OF RANGE (Pacing current out of tolerance; param: high-16 = selected current, low-16 is actual _ _ _ _ _ _ current.)
- a017Error codeTH ERROR DEFIB (Defib. error report) _ _
- a018Error codeTH ERROR PACER (Pacer error report) _ _