Physio-Control All models error codes
Codes
- 000aError codeUT ERROR DAC FAILURE (ECG DAC self-test failed. ECG DAC failure after cold boot.) _ _ _
- 000cError codeUT ERROR ADC TEST REG (ADC Test Register Failure. ADC Test Register test failure. Failure to read the register after _ _ _ _ three tries. May also be caused by the serial channel not responding.)
- 000dError codeUT ERROR ADC CAL NOT COMPLETE (ADC busy bit not clear 150 ms after calibration. ADC Self-Test Calibration test _ _ _ _ _ failure.)
- 000eError codeUT ERROR VP FLASH ID UNKNOWN (Unknown manufacture/device ID for voice/printer flash.) _ _ _ _ _
- 000fError codeUT ERROR DP FLASH ID UNKNOWN (Unknown manufacture/device ID for data/program flash.) _ _ _ _ _
- 1Error codeService LED - Illuminates when service error codes are written into the Service Log (accessed in the Service/Status menu, see Displaying the Service/Status Submenu (p. 106)). See Troubleshooting (p. 98) for information about error codes
- 1fError codeUI ERROR EXTRA INFORMATION (Extra error code information for an above error.) _ _ _
- 0002Error codeUT ERROR FLASH VPP (Error during flash block erase. Valid for all flash sizes.) _ _ _
- 2Error codeSpO2/SpCO/SpMet sensor cable port (optional) - Connection port for the pulse oximeter, which noninvasively checks the saturation of oxygen, carboxyhemoglobin concentration, and methemoglobin concentration in arterial blood
- 0003Error codeUT ERROR FLASH ERASE (Flash memory block erase failure. Valid for all flash sizes.) _ _ _
- 3Error codeNIBP pneumatic tubing port (optional) - Port for connection to the blood pressure tubing which connects to the cuff. NIBP measures the blood pressure of the adult or pediatric patient
- 3fError codeDM ERROR EXTRA INFORMATION (Extra error code information for an above error.) _ _ _
- 0004Error codeUT ERROR FLASH 8BIT WRITE (Error during 8-bit flash write. Error status bits indicate error information. Flash not _ _ _ _ updated.)
- 0005Error codeUT ERROR FLASH 16BIT WRITE (Error during 16-bit flash write. Error status bits indicate error information. Flash not _ _ _ _ updated.)
- 0006Error codeUT ERROR FLASH PAGE WRITE (Error during 16-bit flash write. Error status bits indicate error information. Flash not _ _ _ _ updated.)
- 0008Error codeUT ERROR ADC READ (Error during ADC read. ADC serial channel not available.) _ _ _
- 10Error codeALARMS button and LED - Open alarms menu or silences alarms. The LED is illuminated when alarms are enabled and flashes when an alarm condition occurs
- 11Error codeOPTIONS button - Accesses optional functions. The options menu selections are: PATIENT, PACING, DATE/TIME, ALARM VOLUME, ARCHIVES, PRINT, and USER TEST
- 12Error codeInappropriate screen message response
- 13Error codePrinter Error: a. Possible A12 Printer Assembly failure. b. Verify that the appropriate connecting cables and wire harnesses are functional
- 15Error codePacer out of calibration: a. Complete the TCP-Pacer Self-Calibration procedure, and then conduct Performance Inspection Procedures. b. Verify that the appropriate connecting cables and wire harnesses are functional. c. Possible A04 Therapy PCB failure
- 17Error codeBiphasic fault or unknown energy: a. Verify that the appropriate connecting cables and wire harnesses are functional. b. Possible A22 Biphasic PCB failure. c. Possible A04 Therapy PCB failure
- 18Error codePACER button and LED - Activates pacer function. LED illuminated when function is activated and flashes with each current pulse
- 19Error codeDump error code (a02c): a. Verify that the appropriate connecting cables and wire harnesses are functional. b. Possible A04 Therapy PCB failure. c. Possible A22 Biphasic PCB failure. d. Possible A01 System PCB failure
- 20Error codeCURRENT button - Increases or decreases pacing current. The up or down arrows on button adjusts the pacing current in 10 mA increments, or rotate the SPEED DIAL to change the current in 5 mA increments
- 28Error codeThe current device software version is recommended, contact Stryker Field Service (see Service Information (p. 19))
- 100aError codeUI ERROR NIBP KEY SEEN (This unit is not configured to support NIBP, but the software saw a key closure of this key.) _ _ _ _
- 100bError codeUI ERROR CURRENTUP KEY SEEN (This unit is not configured to support pacing, but the software saw a key closure of _ _ _ _ this key.)
- 100cError codeUI ERROR CURRENTDOWN KEY SEEN (This unit is not configured to support pacing, but the software saw a key _ _ _ _ closure of this key.)
- 100dError codeUI ERROR RATEUP KEY SEEN (This unit is not configured to support pacing, but the software saw a key closure of this _ _ _ _ key.)
- 100eError codeUI ERROR RATEDOWN KEY SEEN (This unit is not configured to support pacing, but the software saw a key closure of _ _ _ _ this key.)
- 100fError codeUI ERROR PACER KEY SEEN (This unit is not configured to support pacing, but the software saw a key closure of this _ _ _ _ key.)
- 103bError codeUI ERROR MISSING LANGUAGE (Configured language was not available.) _ _ _
- 400aError codeSM ERROR BAD CRC (CRC in program flash bad. Program test failure; value = high 16 bits expected CRC, low 16 bits _ _ _ include computed CRC.)
- 400bError codeSM ERROR CRC FAILURE (Program contents failed CRC test; value = high 16 bits expected CRC, low 16 bits include _ _ _ computed CRC).)
- 400cError codeSM ERROR VOLTAGE LOW (ADC voltage reading low. HW voltage low; status code = high 8 bits contains ADC value, _ _ _ low 8 bits contains channel #.)
- 400dError codeSM ERROR VOLTAGE HIGH (ADC voltage reading high. HW voltage high; status code = high 8 bits contains ADC value, _ _ _ low 8 bits contains channel #.)
- 1005Error codeUI ERROR DISPLAY SELF TEST (Self-test failed. Upper 16 bits of status code contain the expected CRC; lower 16 bits _ _ _ _ contain the actual CRC.)
- 1006Error codeUI ERROR ENERGY FAULT (Defib charge out of 15% tolerance. Occurs only during manual mode.) _ _ _
- 1007Error codeUI ERROR 12LEAD KEY SEEN (This unit is not configured to support 12-lead, but the software saw a key closure of this _ _ _ _ key.)
- 1008Error codeUI ERROR ANALYZE KEY SEEN (This unit is not configured to support AED mode, but the software saw a key closure of _ _ _ _ this key.)
- 1010Error codeUI ERROR PAUSE KEY SEEN (This unit is not configured to support pacing, but the software saw a key closure of this _ _ _ _ key.)
- 1037Error codeUI ERROR UNCONFIGURED BUTTON (Button test detected key that should not be present in current hardware _ _ _ configuration.)
- 1038Error codeUI ERROR CPR KEY SEEN (This unit is not configured to support AED mode, but the software saw a key closure of this _ _ _ _ key.)
- 3005Error codeDM ERROR DATABASE ERASE ADJUST (Not able to write new lines for new oldest record; disables flash.) _ _ _ _
- 3006Error codeDM ERROR DATABASE ERASE FAILED (Erase block failed; disables flash; param = block requested.) _ _ _ _
- 3007Error codeDM ERROR DATABASE ERASE VERIFY (Verification of erased block failed; disabled flash; param = block.) _ _ _ _
- 3008Error codeDM ERROR DATABASE FLASH ERASE (Erase database failed; disables flash; param = block # of failure.) _ _ _ _
- 4009Error codeSM ERROR RAM FAILURE (RAM failure during self-test. 16-bit ram test failure; param = address of failure.) _ _ _
- 4010Error codeSM ERROR SERVICE LED (Service LED failure. LED expected to be on, but it is not.) _ _ _