- of5Error codeAll features of 5.03.00 version, and corrects translation errors in French, German, and Spanish languages. This is the current release for English, French, German, and Spanish pumps
- or12VError codeMAINS LED does not light when plugged into a known good AC or 12VDC outlet. BATTERY LED lights when pump is powered up
- OTH01Error codeOther failures
- P0Error codechecksum error on get prev syserr() _ _
- P2Error codechecksum error on event log only in start logs() _
- P3Error codechecksum error on both logs in start logs() _
- P8Error codePump Module power
- P15Error codeLEDs/Icons PCB
- PR01Error codeIncorrect soft Bootstrap CRC
- PR02Error codeIncorrect program CRC
- PR03Error codeA certain program task not responding
- PS02Error codePusher drive travel deviated from expected position based upon motor control
- PS03Error codeIncorrect CRC of the pusher position table
- PS04Error codePusher position sensor readings out of range for the inserted syringe
- PS05Error codePusher position sensor has failed
- PS06Error codeInvalid pusher position codes
- R118Error codeR118 on the UIM PCB is burnt (underside, immediately below U66 and U4)
- ror32Error codeSystem Error 32
- ror33Error codeSystem Error 33
- RT02Error codeReal time clock software module has failed
- SC03Error codeSound control software module has failed
- SF01Error codeSyringe force sensor reading out of range
- SF02Error codeSyringe force test reading out of range
- SF03Error codeSyringe force 0 reference reading out of range
- SF04Error codeIncorrect CRC of the force (pressure) table
- SF05Error codeSyringe force sensor reading not change
- SF06Error codeSyringe force reference voltage drift
- SF07Error codeIncorrect occlusion pressure level value in EEPROM
- SM01Error codeStepper motor software module has failed
- SM02Error codeMomentarily stepper motor rotation speed is high in comparison of inputs steps and encoder
- SM03Error codeMomentarily stepper motor rotation speed is low in comparison of input steps and encoder, or is not rotating when it should
- SM04Error codeAccumulated stepper motor rotation speed is low or high in comparison of input steps and encoder
- SR01Error codeSRAM test failed
- SRE01Error codeSyringe gripper and push lever (retainer) sensors readings out of range (>1.4V)
- SRE02Error codePusher lever or drive engaged sensors are damaged
- SRE03Error codeSyringe retainer (pusher lever, syringe gripper) or drive engaged sensors are damaged
- SS01Error codeSyringe size sensor readings out of range
- SS02Error codeIncorrect CRC of the syringe table in EEPROM
- SS03Error codeIncorrect CRC of the syringe table in SRAM
- to20Error codeTesting downstream occlusion pressure
- to100Error codeCollecting pumped solution during volume delivery accuracy tests
- TR01Error codeInternal microcontroller TRAP activated
- TR02Error codeInternal microcontroller TRAP activated
- TR03Error codeInternal microcontroller TRAP activated
- TR04Error codeInternal microcontroller TRAP activated
- UL-2601Error code< 300 µA earth leakage with ground open < 100 µA earth leakage with ground intact
- WD05Error codeW-D circuit failed
- WD06Error codeW-D circuit failed
- WD07Error codeWatch-Dog circuit activated
- WD08Error codeWatch-Dog circuit activated