Error code
Baxter All models code PS04
No approved translation for this record yet; showing the original text.
Meaning
Pusher position sensor readings out of range for the inserted syringe
Member details
Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.
View details as a memberPossible causes
Calibrate pusher position and syringe size sensors. If the problem persists, check and replace encoder ruler, sensors unit B6660032 or syringe size sensor if necessary
Solution
Solution steps are available to members and are not shown on this page. Answer the questions to open the solution.