Skip to content
RemFixitSearch fault code

Error code

Baxter All models code SF02

No approved translation for this record yet; showing the original text.

Meaning

Syringe force test reading out of range

Member details

Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.

View details as a member

Possible causes

Calibrate syringe force sensor. If the problem persists: Check syringe force sensor and replace syringe pusher with syringe force sensor if necessary. Check syringe force sensor amplifier circuit and replace Drive Unit board if necessary

Solution

Solution steps are available to members and are not shown on this page. Answer the questions to open the solution.

Remote technician support

Get help from a technician over a video call.