Stryker defibrilatör All models error codes
Codes
- a10cError codeDE ERROR CHG ENABLE (CHG EN1 is stuck high.) _ _ _ _
- a10dError codeDE ERROR DUMP ENERGY (Defib HW error.) _ _ _
- a10eError codeDE ERROR RCV (Defib HW error.) _ _
- a11aError codeDE ERROR TEST CPU (Defib HW error.) _ _ _
- a11dError codeDE ERROR XFER TIMEOUT (Defib HW error.) _ _ _
- a11eError codeDE ERROR BUTTONS UP (Defib HW error.) _ _ _
- a11fError codeDE ERROR SYNC INTERRUPT (Defib HW error.) _ _ _
- a12aError codeDE ERROR XFER PADDLE (Defib HW error.) _ _ _
- a12cError codeDE ERROR CHG INHIBIT (Defib HW error.) _ _ _
- a12dError codeDE ERROR CHG ENABLE FAIL (Charge enable feedback indicates not charging.) _ _ _ _
- a12eError codeDE ERROR BTE FAULT (Cedar BTE Fault Line State.) _ _ _
- a12fError codeDE ERROR BTE FAULT CLEARED (Cedar BTE Fault Line State.) _ _ _ _
- a20dError codePA ERROR PACE OVERRUN (Pacing pulse process overrun. Set current to zero.) _ _ _
- a20eError codePA ERROR PULSE WIDTH (Pacing pulse too long.) _ _ _
- a20fError codePA ERROR A2D INT (Internal A/D conversion timeout. Set current to zero.) _ _ _
- a111Error codeDE ERROR ENERGY NOT ZERO (Unexpected energy while biphasic is in high-pot test.) _ _ _ _
- a112Error codeDE ERROR TEST XFER ENABLE (Problem with transfer; turn on at main.) _ _ _ _
- a113Error codeDE ERROR TEST XFER ENGAGE (Error reading the transfer engage feedback.) _ _ _ _
- a114Error codeDE ERROR TEST DUMP RELAY (Biphasic error.) _ _ _ _
- a115Error codeDE ERROR TEST ADC (Defib HW error.) _ _ _
- a116Error codeDE ERROR TEST DAC (Defib HW error.) _ _ _
- a117Error codeDE ERROR TEST HARDWARE (Defib HW error.) _ _ _
- a118Error codeDE ERROR TEST RAM (Defib HW error.) _ _ _
- a119Error codeDE ERROR TEST ROM (Defib HW error.) _ _ _
- a120Error codeDE ERROR SELF TEST INCOMPLETE (Defib HW error.) _ _ _ _
- a123Error codeDE ERROR CAL RCV CRC (CRC failed for new calibration data.) _ _ _ _
- a124Error codeDE ERROR CAL NVM CRC (Cannot write energy table to EEPROM.) _ _ _ _
- a125Error codeDE ERROR DAC WRITE (Byte could not be written to the DAC through the SPI interface.) _ _ _
- a126Error codeDE ERROR ADC READ (Cannot read from ADC.) _ _ _
- a127Error codeDE ERROR TEST MODE (Must be idle to switch to test mode.) _ _ _
- a129Error codeDE ERROR XFER CABLE (Defib HW error.) _ _ _
- a130Error codeDE ERROR BTE RESET (Defib HW error.) _ _ _
- a131Error codeDE ERROR NO BTE HW (Biphasic HW not found.) _ _ _ _
- a132Error codeDE ERROR NO BTE XFER (Defib HW error.) _ _ _ _
- a133Error codeDE ERROR BTE CEDAR DRV HI (A/D high bit sticky.) _ _ _ _ _
- a201Error codePA ERROR LONG WDT (Long watchdog timer test failed.) _ _ _
- a202Error codePA ERROR SHORT WDT (Short watchdog timer test failed.) _ _ _
- a203Error codePA ERROR SCI RCV (SCI received overrun, framing or parity. Unit used near high EMI causing these issues.) _ _ _
- a204Error codePA ERROR 2MS OVERRUN (2 ms ECG sampling overrun.) _ _ _
- a208Error codePA ERROR MSG RESYNC (Received message incomplete.) _ _ _
- a209Error codePA ERROR MSG SIZE (Received msg size error/input buff full.) _ _ _
- a210Error codePA ERROR A2D EX (External A/D conversion timeout. Current set to zero.) _ _ _
- a211Error codePA ERROR SPI (SPI transfer timeout. Current set to zero.) _ _
- a212Error codePA ERROR RAM TEST (RAM test failed. Reset Pacer Processor.) _ _ _
- a213Error codePA ERROR ROM TEST (ROM CRC test failed.) _ _ _
- a214Error codePA ERROR CPU TEST (Stack overrun occurred.) _ _ _
- a215Error codePA ERROR STACK CHECK (Isolated +5 V ref. out of range.) _ _ _
- a216Error codePA ERROR V ISO MON (HV present when not pacing.) _ _ _ _
- a217Error codePA ERROR V 12V MON (+12 V voltage out of range.) _ _ _ _
- a218Error codePA ERROR V HVIS SENSE (HV present when not pacing.) _ _ _ _