Stryker defibrilatör All models error codes
Codes
- a5fError codeTH ED ERROR UNKNOWN (Energy Delivery Hardware failure (ED reported an unknown fault ID)) _ _ _
- a50aError codeTH ED ERROR MSG CRC (Energy Delivery Comm failure (ED received 3 consecutive messages with CRC failures)) _ _ _ _
- a50cError codeTH ED ERROR SEQUENCING (Energy Delivery Comm failure (ED received 3 consecutive messages out of sequence)) _ _ _
- a50fError codeTH ED ERROR CAP OVER VOLTAGE (Energy Delivery Hardware failure (Cap voltage exceeds maximum)) _ _ _ _ _
- a51aError codeTH ED ERROR DEFIB CAL (Energy Delivery Hardware failure (Energy calibration table CRC failure, ED will use default table)) _ _ _ _
- a51bError codeTH ED ERROR ADC COMPARE (Energy Delivery Hardware failure (VCAP1 and VCAP2 do not match)) _ _ _ _
- a51cError codeTH ED ERROR VOLTAGE ON CAP (Energy Delivery Hardware failure (Voltage on the CAP when not expected)) _ _ _ _ _
- a51eError codeTH ED ERROR RELAY TEST (Energy Delivery/System Control Hardware failure) _ _ _ _ (1=SYS XFER EN line not high or ED UP XFR RLY EN line not low _ _ _ _ _ _ 2=SYS XFER EN line not low or ED UP XFR RLY EN line not low _ _ _ _ _ _ 3=SYS XFER EN line not low or ED UP XFR RLY EN line not high) _ _ _ _ _ _
- a51fError codeTH ED ERROR OVER VOLTAGE RESET _ _ _ _ _
- a507Error codeTH ED ERROR OSCILLATOR (Energy Delivery Hardware failure (ED oscillator fault, ED switches to the internal DCO clock)) _ _ _
- a508Error codeTH ED ERROR TEST XFER OPEN (Energy Delivery Hardware failure (XFER Engage signal is open when expected to be _ _ _ _ _ closed))
- a509Error codeTH ED ERROR TEST XFER CLOSE (Energy Delivery Hardware failure (XFER Engage signal is closed when expected to be _ _ _ _ _ open))
- a510Error codeTH ED ERROR CHARGING TIMEOUT (Energy Delivery Hardware failure (Cap stays CHARGED for 65 seconds)) _ _ _ _
- a511Error codeTH ED DUMP TIMEOUT (Energy Delivery Hardware failure (Dump Timeout - Capacitor energy dump to less than 1 joule takes _ _ _ longer than 25 seconds))
- a513Error codeTH ED ERROR PACE PULSE WIDTH (Energy Delivery Hardware failure (Pace pulse width greater than 30ms)) _ _ _ _ _
- a514Error codeTH ED ERROR PACE CHARGE TIMEOUT (Energy Delivery Hardware failure (Charge timeout during pacing)) _ _ _ _ _
- a518Error codeTH ED ERROR UNEXPECTED WATCHDOG (Energy Delivery Hardware failure (Unexpected watchdog reset)) _ _ _ _
- a519Error codeTH ED ERROR HTEST (Energy Delivery Hardware failure (H-Bridge test failure)) _ _ _
- b001Error codePR ERROR TEMP TOO LOW (Printhead ADC reading too low; param = ADC value. May be associated with _ _ _ _ UT ERROR DAC FAILURE and UT ERROR ADC READ.) _ _ _ _ _ _
- b002Error codePR ERROR TEMP TOO HIGH (Printhead ADC reading too high; param = ADC value. May be associated with _ _ _ _ UT ERROR DAC FAILURE and UT ERROR ADC READ.) _ _ _ _ _ _
- b003Error codePR ERROR SELF TEST FAIL (Invalid CRC read from HW; param = ADC value.) _ _ _ _
- b004Error codePR ERROR WRONG PRINTER TYPE (Saw incorrect printer for config; 50/100 mismatch.) _ _ _ _
- by1Error codeAfter a shock is delivered you will see and hear START CPR. A countdown timer (min:sec format) continues for the duration specified in the CPR TIME 1 setup option. Note: If the CPR metronome is set to be on during CPR times, you will hear audible “tocks” and ventilation prompts or tones
- c008Error codePM ERROR LOST COMMS (Lost communications with power PCB.) _ _ _
- c009Error codePM ERROR CLOCK FAILURE (Power PCB is running on its internal oscillator block because it detected an external clock failure.) _ _ _
- CO2Error codeCO2 fails calibration
- L05Error codeSPACER - FOAM, SYS PCB CAPS
- M19Error codeASSY - MOUNTING, BRACKET, SYSTEM, THERAPY PCB
- M30Error codeGASKET - CONNECTOR, INVASIVE PRESSURE
- M49Error codeSTRAP - GROUND, BTE PCB
- M54Error codeDIELECTRIC SHIELD, SYSTEM PCBA
- SPR11Error codeGrup 1 Sınıf B
- to6Error codeCO2 displays XXX in the CO2 region of the display with Service LED OFF
- W01Error codeW01 - CABLE ASSY - POWER, SYSTEM, PCBA
- W04Error codeW04 - CABLE ASSY - FLEX, SYSTEM PCB/ INTERFACE PCB
- W06Error codeW06 - CABLE ASSY - BACKLIGHT INV, INTERFACE PCB
- W12Error codeW12 - CABLE ASSY - PRINTER CONTROL KEYPAD, INTERFACE PCB
- W13Error codeW13 - CABLE-ASSY, MAIN KEYPAD, INTERFACE PCB
- W16Error codeW16 - CABLE ASSY, RIBBON, INTERFACE PCB/CHART RCDR
- W18Error codeW18 - CABLE ASSY-FLEX, LCD, INTERFACE PCB
- W21Error codeW21 - CABLE, ASSY - FLEX, Sp02, OEM, PCB, ROHS
- W26Error codeW26 - CABLE ASSY - CO2 MODULE, OEM PCB
- W27Error codeW27 - CABLE ASSY-RIBBON, NIBP/OEM PCB
- W31Error codeASSY, CABLE, CO2 MODULE TO PCBA, LP15
- W33Error codeW33 - WIRE HARNESS- INVASIVE PRESSURE 1- 2, ROHS
- W35Error codeW35 - CONNECTOR CABLE ASSEMBLY - TEMPERATURE, ROHS