Stryker defibrilatör All models error codes
Codes
- a00aError codeTH ERROR DE WRONG ENG SELECT (Incorrect energy selected; param = energy index.) _ _ _ _ _
- a00bError codeTH ERROR SHOCK NOT DELIVERED (Shock not delivered.) _ _ _ _
- a00dError codeTH ERROR CHARGING EXPIRED (Charging time expired.) _ _ _
- a00fError codeTH ERROR CAP OUT OF RANGE (Capacitor is out of range. Calibration failure.) _ _ _ _ _
- A01Error codeSystem/Interface PCB Cable (W04) Replacement
- a01aError codeTH ERROR PACER FAULT (Pacing fault condition occurred [rate(0), current(1), pulse width(2)], limit exceeded; param = _ _ _ pacer-fault type.)
- a01bError codeTH ERROR DEFIB WDT DISABLE FAIL (Unable to turn off defib WDT; param = ASIC defib ctrl register.) _ _ _ _ _
- a002Error codeTH ERROR DEFIB LINK DOWN (Lost contact with defib processor. Serial communications link between the main and defib _ _ _ _ processor is not functioning.This unit can no longer administer defib therapy. Cycling power may clear the error temporarily, but unit is questionable; param = last defib message time stamp.)
- a02bError codeTH ERROR DEFIB CONFIG (Conflicting defib type. Cold boot if test bit set; otherwise, disable biphasic; param = test-bit _ _ _ setting.)
- a02cError codeTH ERROR DUMP LINE FAIL (One of dump lines failed; param = test ID.) _ _ _ _
- a02dError codeTH ERROR WRONG DEFIB TYPE (Wrong Defib. type for software.) _ _ _ _
- a02eError codeTH ERROR ADC READ (ADC read failure when getting cap charge.) _ _ _
- a02fError codeTH ERROR MSG ID UNKNOWN (Message ID from System PCB is unknown) _ _ _ _
- a003Error codeTH ERROR PACER LINK DOWN (Lost contact with pacer processor. Serial communications link between the main and pacer _ _ _ _ processor is not functioning.)
- A03Error codePCB ASSY - POWER, LP15
- a004Error codeTH ERROR UNEXPECTED ENERGY (Unexpected energy in the capacitor. No charge delivered.) _ _ _
- A04Error codeA04 - PCB ASSY - THERAPY, BIPHASIC
- a005Error codeTH ERROR CAP OVERCHARGED (Over-charged capacitor.) _ _ _
- A05Error codePCB ASSY - INTERFACE, LP15 V4
- A06Error codeA06 - PCB ASSY - OEM, LP15
- a008Error codeTH ERROR DEFIB DISABLE (No communication with defib HW.) _ _ _
- a010Error codeTH ERROR PA RATE OUT OF RANGE (Pacing rate out of tolerance; param: high-16 = pacer selected range, low-16 = pacer _ _ _ _ _ _ actual rate.)
- a10bError codeDE ERROR CHG TIME (Takes too long to reach charge.) _ _ _
- a011Error codeTH ERROR PA CURRENT OUT OF RANGE (Pacing current out of tolerance; param: high-16 = selected current, low-16 is _ _ _ _ _ _ actual current.)
- a017Error codeTH ERROR DEFIB (Defib. error report) _ _
- a018Error codeTH ERROR PACER (Pacer error report) _ _
- a020Error codeTH ERROR PACER DISABLE (Pacer disabled; cannot communicate with processor.) _ _ _
- a021Error codeTH ERROR CAP CHARGE FAIL (Cap. stays zero while charging. No charge; param = defib setting.) _ _ _ _
- a022Error codeTH ERROR CORRUPT ENERGY SELECT (Energy select corrupt; param = main energy selection.) _ _ _ _
- a023Error codeTH ERROR XFER ENABLE ON (Defib transfer-enable line high unexpected. Defib transfer-enable line was not off during _ _ _ _ startup.)
- a024Error codeTH ERROR VCAP2 SATURATED (VCAP2 reading full scale all the time. Reading not processed.) _ _ _
- a026Error codeTH ERROR ENERGY RESIDUE (There is still energy on the cap.) _ _ _
- a027Error codeTH ERROR PA RATE CORRUPT (Pacing rate storage corrupted; param: high-16 = rate selected, low-16 = actual rate.) _ _ _ _
- a028Error codeTH ERROR CAL ENERGY FAIL (Calibrated voltage is out of range; param: high-16 = table index, low-16 = voltage count.) _ _ _ _
- a029Error codeTH ERROR BTE FAIL (Error condition with BTE board; param: high-16 bits = fault type, low-16 bits = cedar state.) _ _ _
- a032Error codeTH ERROR PC LINK DOWN (Lost contact with the Paddles Control subsystem) _ _ _ _
- a033Error codeTH ERROR PS LINK DOWN (Lost contact with the Paddles Sensing subsystem) _ _ _ _
- a034Error codeTH ERROR ED LINK DOWN (Lost contact with the Energy Delivery subsystem) _ _ _ _
- a035Error codeTH ERROR COMPATIBILITY ERROR (Wrong version of software is installed) _ _ _
- a036Error codeTH ERROR NO SELFTEST RESPONSE (Did not respond to 3AM or User Test) _ _ _ _
- a037Error codeTH ERROR XFER ENGAGE (Transfer was not engaged when expected) _ _ _
- a038Error codeTH ERROR VCOMP IMPED INVALID (Error indicates that a valid normal patient impedance was not received within specified _ _ _ _ time period, and held off charge will begin without voltage compensation)
- a039Error codeTH ERROR INVALID STATE (If the therapy processor (PC, PS, or ED) reports an incorrect state after 3 attempts to correct, the _ _ _ therapy processor will be reset.)
- a101Error codeDE ERROR LONG WDT (Long watchdog test failed or watchdog did not reset in time.) _ _ _
- a102Error codeDE ERROR SHORT WDT (Short watchdog test failed.) _ _ _
- a103Error codeDE ERROR SCI RCV (Serial port receiver error.) _ _ _
- a104Error codeDE ERROR XFER KEY (Defib HW error.) _ _ _
- a106Error codeDE ERROR ENERGY OUT OF BOUND (VCAP-1 over/under charge.) _ _ _ _ _
- a107Error codeDE ERROR HP ENG SELECT (Cannot determine the rotary setting.) _ _ _ _
- a109Error codeDE ERROR CAL CRC (Calibration Table CRC error.) _ _ _