Philips Trilogy taşınabilir ventilatör All models error codes
Codes
- E-9Error code• Motor Capacitor not installed • Motor Connector not soldered properly • Motor Rotor Locked or other mechanical problem
- E-53Error code• The measured Exhaled Tidal Volume is greater than or equal to the alarm setting • Flow Sensor problem • Active exhalation valve problem
- E-54Error code• The measured Exhaled Tidal Volume is less than or equal to the alarm setting • High leak • Flow sensor problem • Active exhalation valve problem
- E-67Error code• The sensor has drifted so much that the positive or negative displacement of the zero point has moved the maximum or minimum calculable pressure value up or down somewhat. Enough positive displacement can reduce the top end of pressure readings below that expected by the Trilogy design. By similar reasoning, enough negative displacement can result in a condition where the pressure sensor cannot read low enough for Trilogy • The drift data for a sensor is also collected at zero pressure, and is converted to engineering units. When the absolute value (since the sensor can drift in either direction) of this drift exceeds a configured limit, a range error has occurred
- E-68Error code• Error indicates a short in the Li-Ion battery discharge path on the Power Management PCA. Normally, the Host software recognizes a useable Detachable and Internal battery and attempts to disable the discharge path of the Internal Battery. Since the Internal Battery discharge path has failed shorted, current will be shared by the Internal and Detachable Battery
- E-69Error code• The sensor has drifted so much that the positive or negative displacement of the zero point has moved the maximum or minimum calculable pressure value up or down somewhat. Enough positive displacement can reduce the top end of pressure readings below that expected by the Trilogy design. By similar reasoning, enough negative displacement can result in a condition where the pressure sensor cannot read low enough for Trilogy • The drift data for a sensor is also collected at zero pressure, and is converted to engineering units. When the absolute value (since the sensor can drift in either direction) of this drift exceeds a configured limit, a range error has occurred
- E-70Error code• The sensor has drifted so much that the positive or negative displacement of the zero point has moved the maximum or minimum calculable pressure value up or down somewhat. Enough positive displacement can reduce the top end of pressure readings below that expected by the Trilogy design. By similar reasoning, enough negative displacement can result in a condition where the pressure sensor cannot read low enough for Trilogy • The drift data for a sensor is also collected at zero pressure, and is converted to engineering units. When the absolute value (since the sensor can drift in either direction) of this drift exceeds a configured limit, a range error as occurred
- E-76Error code• Defective EEPROM
- E-77Error code• Defective EEPROM • Program Execution Error
- E-80Error code• The sensor has drifted so much that the positive or negative displacement of the zero point has moved the maximum or minimum calculable pressure value up or down somewhat. Enough positive displacement can reduce the top end of pressure readings below that expected by the Trilogy design. By similar reasoning, enough negative displacement can result in a condition where the pressure sensor cannot read low enough for Trilogy • The drift data for a sensor is also collected at zero pressure, and is converted to engineering units. When the absolute value (since the sensor can drift in either direction) of this drift exceeds a configured limit, a range error has occurred
- E-81Error code• Faulty SD Card
- E-84Error codeCard Error
- E-85Error codeCard Error
- E-86Error codeCard Error
- E-87Error codeCard Error
- E-88Error codeCard Error
- E-89Error codeCard Error
- E-94Error code• Barometric pressure sensor is reading maximum or minimum counts for an extended period of time
- E-323Error code• When the device encounters errors while it has started to apply the Rx Setting from the SD Card, the device stops the Rx update and tries to restore back the old setting on the device. During the restoration of these old settings of the device, if there is an error this error code is generated
- E-333Error code• U24 circuitry failure • U26 circuitry failure
- E-339Error codePressure Sensor Drift - turn off for 20 secs
- E-342Error codeHigh Priority
- E-343Error codeHigh Priority
- E-344Error code• Failure of Oxygen Blending Module
- E-345Error code• Internal O sensor at end of life or watchdog failure 2
- E-349Error code• Communication between Host and OBM was lost
- E-351Error codeLow Oxygen Flow
- E-352Error codeHigh Oxygen Flow
- E-354Error code• Unit not calibrated • O Sensor Calibration Table corrupted 2
- E-356Error code• Before a drift is computed for this sensor, its auto-null valve is opened to atmosphere. A test is performed to ensure that the valve has fully opened in accordance with the time provided by the technical specification
- E-357Error code• Control Flow Sensor out of electrical spec
- E-358Error code• Proximal Flow Sensor out of electrical spec
- E-359Error codeHigh Internal Oxygen
- E-360Error codeLow Vte
- O2Error code• Verify seating of the O 2 proportional valve • Verify connection of green tube to manifold • Verify green tube to port clip adaptor • Verify O-rings on port clip adapter to pressure sensor are present • Pressure sensor leaking
- ogy202Error codeIt occurs when the breathing circuit is disconnected or has a large leak. The alarm will automatically terminate when the leak is fixed for 6 seconds and breaths are delivered again
- SPR11Error codeClass B