Philips Trilogy taşınabilir ventilatör All models error codes
Codes
- E-21Error code• Problem with serial communication between DSP and Host
- E-22Error code• Motor Temperature sensor failed • Problem with U37 circuitry
- E-31Error code• Problem with +3.3V DSP _ • U23 failure • DSP failure
- E-32Error code• Faulty Sensor Board Cable • Faulty Sensor Board Circuitry
- E-180Error codeExternal Batt Disconnected
- E-181Error codeDetach Batt Disconnected
- E-189Error code• Detachable or Internal Li-Ion usable and V battLi _ < 9 volts for greater than 10 sec indicates Li-Ion battery unable to power the System • PMB fault
- E-191Error code• AFE chip does not receive the appropriate frequency on the WDI pin from the Gas Gauge chip. Both of these chips are i the Li-Ion battery pack. • Internal Li-Ion fault
- E-192Error code
- E-193Error code• Internal Li-Ion battery permanent failure
- E-194Error codeReplace Detachable Battery
- E-195Error code• Internal Li-Ion Battery State of Health < 50%. • Full charge capacity (FCC) is less than 51% of the Design Capacity • Internal Li-Ion FCC error • Internal Li-Ion fault • Internal Li-Ion end of life
- E-196Error codeReplace Detachable Battery
- E-197Error code• Internal Li-Ion not present as indicated by the Charger Chip status • PMB fault • Internal Li-Ion battery pack fault
- E-198Error code• Circuitry failure
- E-199Error code• Detachable Battery harness failure • Power Management Board Failure • Circuitry failure
- E-200Error code• Circuitry failure
- E-201Error code• Power Management Board Failure • Circuitry failure
- E-202Error code• Circuitry failure
- E-203Error code• Circuitry failure
- E-204Error code• U3 failure • CPLD failure
- E-205Error code• Error indicates a shorted Q25, Q26 in the battery discharge path on the Power Management PCA.
- E-207Error code• Q11 failure on Power Management PCA
- E-218Error codeBatt Discharge Stopped-Temp
- E-219Error codeBatt Discharge Stopped-Temp
- E-230Error codeBattery Not Charging - Temp
- E-231Error codeBattery Not Charging - Temp
- E-237Error codeUpgrade Failed Screen
- E-240Error codeUpgrade Failed Screen
- E-246Error codeInternal Battery Not Charging
- E-247Error codeDetach Battery Not Charging
- E-253Error code• Communication failure • The Host CPU is unable to communicate with the DSP
- E-265Error code• Checks if Li-Ion battery is present by attempting to communicate when the Charger CHip safety signal indicates the battery is not present. Safety signal from the Li-Ion may be disconnected or Charger CHip may have a fault so it is not detecting a Li-Ion battery
- E-266Error codeReplace Detachable Battery
- E-269Error code• Error indicates an open VBULK FET. This MOSFET connects the output of the AC/DC supply to the input of the boost converter. Unit will not run on AC power. Will use battery power if available
- E-270Error code• Error indicates an open Q11, which controls the discharge of the Li-Ion batteries post VbattLi. When the Lithium Ion power is selected, this FET is turned on. If Q11 fails open, there will be no Li- Ion battery power to the system
- E-271Error code• Error indicates an open Q13, Q14 in the Lead Acid battery discharge path on the Power Management PCA. When Pb Acid is selected by the hardware or software, VBATTSENSE would be low. If AC is disconnected, there would be no battery backup from the Li-Ion battery
- E-272Error code• Charger chip failure • FET failure
- E-289Error code• One or more “Ventilator Service Recommended” errors are active in the system
- E-290Error code• One or more “Ventilator Service Required” errors are active in the system
- E-299Error codeCard Error
- E-300Error codeCard Error
- E-301Error codeCard Error
- E-302Error code• Voltage reference on internal ADC failed
- E-304Error code• O sensor failed 2
- E-307Error codeHigh Temperature
- E-308Error codeHigh Temperature
- E-314Error codeFailed -Card is read only
- E-317Error codePrescription Change Failed
- E-320Error codePrescription Change Failed