Philips Trilogy taşınabilir ventilatör All models error codes
Codes
- E-15Error code• Hardware current limiting is not working correctly
- E-17Error code• Big Motor Capacitor not installed or wrong value or at the end of life cycle
- E-100Error code• Software error • Bad table written by Production or Field Service Calibration
- E-101Error code• Program Execution Error
- E-103Error code• Control Flow Sensor out of electrical spec
- E-109Error codeHigh Expiratory Pressure
- E-110Error codeLow Inspiratory Pressure
- E-111Error codeHigh Respiratory Rate
- E-112Error codeLow Respiratory Rate
- E-113Error codeHigh Minute Ventilation
- E-114Error codeLow Minute Ventilation
- E-115Error codeHigh Vti
- E-116Error codeLow Vti
- E-118Error codeLow Circuit Leak
- E-120Error codeApnea
- E-121Error codeCircuit Disconnect
- E-123Error code• Error indicates a shorted Q29 in the Li-Ion Battery discharge path on the Power Management PCA. This MOSFET prevents a low mA sneak current from the Internal Battery when the discharge path is disabled by either the Host software or the hardware
- E-125Error code• 3 reboots occurred within 24 hours
- E-130Error code• The sensor has drifted so much that the positive or negative displacement of the zero point has moved the maximum or minimum calculable pressure value up or down somewhat. Enough positive displacement can reduce the top end of pressure readings below that expected by the Trilogy design. By similar reasoning, enough negative displacement can result in a condition where the pressure sensor cannot read low enough for Trilogy • The drift data for a sensor is also collected at zero pressure, and is converted to engineering units. When the absolute value (since the sensor can drift in either direction) of this drift exceeds a configured limit, a range error has occurred
- E-131Error code• Control pressure sensor is reading maximum or minimum counts for an extended period of time
- E-132Error code• Monitor pressure sensor is reading maximum or minimum counts for an extended period of time
- E-133Error code• Proximal pressure sensor is reading maximum or minimum counts for an extended period of time
- E-134Error code• Control flow sensor is reading maximum or minimum counts for an extended period of time
- E-135Error code• Monitor flow sensor is reading maximum or minimum counts for an extended period of time
- E-136Error code• Air stream Temperature 1 sensor is reading maximum or minimum counts for an extended period of time
- E-137Error code• dP Pressure Temperature sensor is reading maximum counts for an extended period of time
- E-138Error code• Barometric Pressure sensor is reading a count outside of its valid range
- E-139Error code• Software Error
- E-141Error code• Motor connector not soldered correctly • Motor Rotor Locked or other mechanical problem
- E-143Error code• Control or Monitoring Pressure sensor has drifted too far away from the other
- E-145Error code• Both control and Monitoring Pressure sensors failed. • ADC failed
- E-146Error code• Both control and Monitoring Pressure sensors failed. • ADC failed
- E-147Error code• The sensor has drifted so much that the positive or negative displacement of the zero point has moved the maximum or minimum calculable pressure value up or down somewhat. Enough positive displacement can reduce the top end of pressure readings below that expected by the Trilogy design. By similar reasoning, enough negative displacement can result in a condition where the pressure sensor cannot read low enough for Trilogy • The drift data for a sensor is also collected at zero pressure, and is converted to engineering units. When the absolute value (since the sensor can drift in either direction) of this drift exceeds a configured limit, a range error has occurred
- E-149Error code• Before a drift is computed for this sensor, its auto- null valve is opened to atmosphere. A test is performed to ensure that the valve has fully opened in accordance with the time provided by the technical specification. This is performed by ensuring that the control pressure has dropped significantly enough after the valve has opened and a de-bounce time has expired. If the pressure reading before the valve was opened was not high enough (due to ventilator therapy settings) to provide a valid criteria regarding a significant drop in pressure, then the check is not performed. When this condition is in effect for a consecutive number of attempted drifts, the error is reported
- E-158Error code• A serious alarm has occurred before six breaths have been taken after blower is turned on. The lack of therapy data causes back up therapy not to be initiated and the ventilator to become completely inoperative. The alarms that would initiate this possible scenario are: both pressure sensors failing, both pressure and flow sensors failing, a failure of non-volatile memory, or a high pressure patient alarm
- E-160Error code• Obstructed intake • Obstructed flow path • Obstructed or disconnected pressure tubing • Faulty Blower
- E-161Error code• Internal Li-Ion has a TDA due to Discharge Overcurrent or Discharge Short Circuit • PMB fault • System Board fault • Internal Li-Ion fault
- E-162Error codeReplace Detachable Battery
- E-163Error code• Control Flow Sensor out of electrical specification
- E-164Error code• Monitor Flow Sensor out of electrical spec
- E-165Error codeLow Expiratory
- E-170Error codeHigh Inspiratory Pressure
- E-171Error codeHigh Inspiratory Pressure
- E-172Error code• Unit is exceeding 60cm of delivered pressure • High pressure sensor reading • Large amount of drift • Pinched/blocked tubing
- E-174Error code• Can’t communicate with Charger Chip • PMB fault • Internal Li-Ion fault • Detachable Li-Ion fault
- E-175Error code• Can’t communicate with Internal Li-Ion battery after five tries and charger is not trying to wake- up charge the battery • Could take up to 210 seconds for alarm to be reported because the Charger is trying to wake- up charge the Internal Li-Ion • PMB fault • Internal Li-Ion fault
- E-176Error codeReplace Detachable Battery
- E-177Error code• Internal Li-Ion battery not Authentic • Fake battery • PMB fault • Internal Li-Ion fault
- E-178Error codeReplace Detachable Battery
- E-179Error codeAC Power Disconnected