Error code
Carbomed karboksiterapi cihazı All models code O2
No approved translation for this record yet; showing the original text.
Meaning
Possible cause: This oxygen concentrator is equipped with an innovative multi-function sensor to monitor the oxygen concentration of the oxygen output. If the average value for oxygen purity is less than 82% for more than 15 seconds, this alarm is activated (only after a minimum of 2 minutes of operation). Counter measures: - Contact your service provider. Checking the alarm function: - Set the volume flow to "maximum", beyond 5 l/min
Member details
Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.
View details as a memberSolution
Solution steps are available to members and are not shown on this page. Answer the questions to open the solution.
Remote technician support
Get help from a technician over a video call.
Address of this page: https://www.remfixit.com/tr/carbomedkarboksiterapicihazi/tum-modeller/o2