Carbomed karboksiterapi cihazı All models error codes
Codes
- E-002Error codePossible cause: The pressure in the reservoir is less than 550 hPa. This system pressure is too low to ensure a stabile oxygen production process. Counter measures: - Check if the volume flow exceeds 5 l/min! - Check if the device inlet filter (hinter der Service lid in the back) is blocked. - Otherwise: Contact your service provider. Checking the alarm function: - Set the volume flow to „maximum“, exceeding 5 l/min
- E-008Error codePossible cause: The gas temperature exceed 60°C. The oxygen concentration cannot be determined. Counter measures: - see “counter measures for temperature alarm” - Contact your service provider. Checking the alarm function: - Block the exhaust air stream. NOTE! If the internal temperature exceeds 84°C, the temperature fuse is activated and shuts off the device. In this case, contact your service provider
- E-010Error codePossible cause: No sensor data is available for 5 seconds. Counter measures: - Contact your service provider. Checking the alarm function: - Not possible - self diagnostics
- E-020Error codePossible cause: Various kinds of microprocessor errors. NOTE! For safety reasons, oxygen concentrating process is stopped. However the compressor keeps on running. Checking the alarm function: - Not possible - self diagnostics
- E001Error codePossible cause: The operating temperature inside the device is too high (> 65°C). Also make sure that the device has a sufficient clearance to other objects (wall, cupboard, etc.). Further information concerning this matter can be found in the section “Maintenance”. NOTE! For protection, the oxygen supply is set after a further waiting period and the error code "nOP" is displayed. However, the compressor continues to run. Checking the alarm function: - Temporary blockage of the exhaust opening. NOTE! If the internal temperature exceeds 84°C, the temperature fuse is activated and shuts off the device. In this case, contact your service provider
- O2Error codePossible cause: This oxygen concentrator is equipped with an innovative multi-function sensor to monitor the oxygen concentration of the oxygen output. If the average value for oxygen purity is less than 82% for more than 15 seconds, this alarm is activated (only after a minimum of 2 minutes of operation). Counter measures: - Contact your service provider. Checking the alarm function: - Set the volume flow to "maximum", beyond 5 l/min