Belmont Medical kan ısıtıcı All models error codes
Codes
- AT50MError code
- of200Error codeWARNING: Excessive or prolonged recirculation may damage red blood cells by exposing them repeatedly to the rollers inside the pump head
- to200Error codeBolus volume can be changed in the Parameters Set-Up screen (Chapter 4, page 25) or by pressing and holding the BOLUS key in the Infuse screen. The new bolus volume will appear in the VOL (volume) status line with the prefix of BOL (bolus). Releasing the Bolus key will start the infusion. Two sets of numbers are displayed within the BOLUS key space. The top number is the bolus value set and the bottom number is the volume pumped, and is counting up from 0 to the volume set on the key. At the end of the bolus volume, the system beeps and returns to the previously selected flow rate if the previous rate was 50 ml/min or lower. If the previous rate was higher than 50 ml/min, the flow rate will be set to 50 ml/min
- UID100Error codeOut of fluid. Bag clamps not fully opened or fully spiked. Tubing in the Fluid out sensor is not seated firmly in the detector, or tubing is stretched or pulls away from the sensor, due to vacuum in the line. Clogged air vent filter or coarse blood filter. Reservoir or recirculate line is obstructed. Detector electronics defective