Error code
Belmont Medical kan ısıtıcı All models code UID100
No approved translation for this record yet; showing the original text.
Meaning
Out of fluid. Bag clamps not fully opened or fully spiked. Tubing in the Fluid out sensor is not seated firmly in the detector, or tubing is stretched or pulls away from the sensor, due to vacuum in the line. Clogged air vent filter or coarse blood filter. Reservoir or recirculate line is obstructed. Detector electronics defective
Member details
Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.
View details as a memberPossible causes
Press MUTE to silence the alarm. If out of fluid, add additional fluid and press REPRIME. Open bag clamp or fully spike the bag. Reseat the tubing in the fluid out detector and make certain that it is seated firmly in the sensor. If the reservoir chamber stays empty during reprime, the air vent filter, on top of the reservoir chamber, may be clogged. In this case, pierce the fluid bag(s) with bag spikes and fully open clamps to allow the air in the reservoir chamber to escape into fluid bag(s) and allow fluid to fill the reservoir chamber. High amounts of particulates in the blood may clog the coarse blood filter in the reservoir chamber. Replace reservoir chamber or disposable if it is clogged. Power down and service the machine if error persists
Solution
Solution steps are available to members and are not shown on this page. Answer the questions to open the solution.