Roche Diagnostics biyokimya analizörü All models error codes
Codes
- A3Error codeThe Cobas Mira service manual contains a comprehensive list of error codes and their corresponding troubleshooting steps. Consult
- A5Error codeRegular calibration is absolutely crucial for the accuracy and reliability of results generated by the Cobas Mira. Failure to calibrate
- A8Error codeFailure to follow the maintenance guidelines may lead to inaccurate test results, increased downtime, premature wear and tear of
- E-111Error codeThe observer ID is invalid. The observer supervising the current OTS test must log in again
- E-304Error codeMeter is too cold or too warm to test correctly. Power meter off and repeat test after 30 minutes at room temperature
- E-402Error codeA test strip error was detected. Check your procedure and repeat test, using a new test strip if necessary
- E-403Error codeA dosing error was detected. Check your procedure and repeat test, using a new test strip
- E-405Error codeA measurement error occurred. Check your procedure and repeat test, using a new test strip
- E-501Error codeThe self-check detected a general error. Repeat the self-check. If problem persists, power meter off and on again
- E-602Error codeGeneral System Error. The meter is defective. No test is possible. Contact your Roche representative
- E-603Error codeTemperature or heater error. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-605Error codeHardware or electronic error. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-606Error codeMeasurement sequence error. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-607Error codeError with measurement module. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-608Error codeA software error was detected. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-702Error codeCommunication error with DMS. Unlock meter from Base Unit and then redock it. If problem persists, power meter off and on
- E-704Error codeICI communication error with DMS. Power meter off and on, and repeat the last procedure
- E-705Error codeCommunication error with measurement module. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-706Error codeCommunication error with test strip barcode scanner. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-708Error codeGeneral communication error. Power meter off and on, and repeat the last procedure. If problem persists, contact your Roche representative
- E-805Error codeThe meter detected a fault with the EEPROM memory. Power meter off and on. If problem persists, contact your Roche representative
- E-806Error codeThe meter detected a fault with the Flash memory. Power meter off and on. If problem persists, contact your Roche representative
- E-807Error codeThe meter detected a general fault with the memory. Power meter off and on. If problem persists, contact your Roche representative
- Q3Error codeWhat should I do if I encounter an error code I don't understand?