Julabo parafin banyosu All models error codes
Codes
- 200-28Error codeTemperature stability
- 200-30Error codeTemperature stability
- A11Error codeCurrent draw CH (at 230 V)
- A12Error codeCurrent draw CH (at 230 V)
- at0Error codeAmbient temperature range
- C-30Error codeTemperature indication
- C-40Error codeTemperature indication
- C0Error codeWorking temperature sensor
- E01Error codeThe device is being operated with no or too little bath fluid or the level is below the minimum level. Top up the bath fluid. A hose has burst (bath fluid level too low because it has been pumped out). Replace the hose and top up the bath fluid
- E03Error codeThe measured temperature is above the set overtemperature limit
- E04Error codeThe measured temperature is below the set under-temperature limit
- E05Error codeThe cable for the working temperature sensor has been interrupted or short-circuited
- E06Error codeDefect of the working or excess temperature sensor. The working and excess temperature sensors report a temperature difference of more than 20 K
- E14Error codeThe cut-out value of the excessive temperature protector is below the defined working temperature . Set the safety temperature to a higher value
- E15Error codeThe external temperature sensor wire is shorted or broken
- E33Error codeThe cable for the overtemperature safety sensor has been broken or short-circuited
- E38Error codeThe setpoint is set to an external temperature sensor, but no signal is present
- E104Error codeThe AD conversion of the AF overtemperature protection device is malfunctioning
- E106Error codeError in the internal temperature sensor
- E107Error codeError in internal temperature mea- surement
- E108Error codeThe self-locking alarm is still active. Switch off the device at the main switch. Wait for approx. 4 seconds and then switch it on again
- E112Error codeError in the internal temperature sensor
- E114Error codeError in internal temperature sens- ing
- E115Error codeError in internal temperature sens- ing
- E118Error codeThe internal AD converter is defective
- E142Error codeDevice started, external temper- ature actual value source set on the display, external temperature reception timeout via CAN bus
- E143Error codeThe set temperature limit for the overtemperature alarm has been exceeded
- E144Error codeThe temperature has fallen below the set limit for the subtempera- ture alarm
- E301Error codeThe pressure in the external tem- perature control circuit exceeded the entered limit value for 1 sec- ond
- E302Error codeThe pressure in the external temperature control circuit has exceeded the entered limit value for 5 seconds
- E401Error codeTemperature sensor evaporator outlet defective (short circuit)
- E402Error codeTemperature sensor evaporator outlet defective (break)
- E413Error codeEvaporation pressure sensor defective (short circuit)
- E414Error codeEvaporation pressure sensor defective (break)
- E417Error codeCondensation pressure sensor defective (short circuit)
- E418Error codeCondensation pressure sensor defective (break)
- E425Error codeError in refrigeration system
- E426Error codeError in refrigeration system
- E427Error codeError in refrigeration system
- E431Error codeMaximum compressor current exceeded
- E1103Error codeLevel detection error
- E1109Error codeThe viscosity of the heat transfer fluid is too high, or circulation is insufficient
- E1217Error codeMains frequency measurement is faulty
- E1218Error codeMains frequency measurement is faulty
- E1302Error codeThe pressure has fallen below the threshold for shutting off the heat- ing output. Motor malfunction or fluid viscos- ity too high
- E1305Error codeThe pump speed is too low. The pump is defective, or the viscosity of the fluid is too high
- E1427Error codeWarning: Error in refrigeration system
- E1431Error codeWarning: No compressor current detected
- E1502Error codeThe pressure in the temperature control circuit exceeds the set warning limit
- E1503Error codeThe pressure in the temperature control circuit has fallen below the set warning limit