Spectra lazer seviye nivo All models error codes
Codes
- E200Error codeThe footswitch cable is not properly con- Ensure that the footswitch is properly nected to the laser console, or the foot- connected to the laser console and that switch cable is defective. the footswitch cable is not damaged. If the condition continues, restart the laser system. If the condition persists, contact your local Lumenis representative
- E701Error codeThe laser detected an over temperature Let the system remain in standby for five condition. minutes. If the error does not clear, turn off the system and wait for 1 minute, then restart. If the condition persists, contact your local Lumenis representative
- F100Error codeThere is an internal shutter failure Contact your local Lumenis representative
- F101Error codeThere is an internal shutter failure Contact your local Lumenis representative
- F204Error codeFootswitch error There is a three second delay from the time the Ready button is pressed and the time the footswitch can be pressed. Wait for the Ready indicator on the console to illuminate before pressing the footswitch. Check footswitch connections and cord. If the condition persists, contact your local Lumenis representative
- F402Error codeDelivery device error First, remove and inspect the delivery device as instructed in your delivery device operator manual; then, verify that the device is properly connected as instructed. If the condition persists, contact your local Lumenis representative
- F503Error codeInternal system operating errors Contact your local Lumenis representative
- F803Error codePower interlock error Turn the laser on and off. If the condition persists, contact your local Lumenis representative
- SPR11Error codeClass A