Error code
Promega Spectrum kapiller elektroforez genetik analiz All models code han200
No approved translation for this record yet; showing the original text.
Meaning
Poor quality data on Spectrum Compact Capillary Cartridge with less than 200 injections
Member details
Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.
View details as a memberPossible causes
Damage to a capillary or capillaries may result in polymer leakage. Check the Spectrum Compact Capillary Cartridge for evidence of polymer leakage (white crystals appearing somewhere along length of capillary). If damage is detected, replace the Spectrum Compact Capillary Cartridge