Haemonetics aferez cihazı All models code Cz1

No approved translation for this record yet; showing the original text.

Meaning

Long Empty • Check effluent tubing for correct placement and occlusions. • Ensure no fluid transfer from waste bag to bowl. • Check blue and red line for occlusions. Explanation: The air detector did not sense air when expected, indicating that the device may have pumped more than the expected volume of fluid from the bowl or that a tubing occlusion could be preventing fluid from emptying as expected

Member details

Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.

View details as a member

Possible causes

Corrective action: 1. Check effluent tubing for correct placement in the effluent line sensor. 2. Check effluent tubing for kinks and occlusions. 3. Ensure that fluid is not transferring from the waste bag to the bowl. NOTE: lf fluid is transferring from the waste bag to the bowl, waste may have reached the RBC bag; the contents of the RBC bag should be returned to the bowl tobe washed again. 4. Check tubing for correct placement in air detector. 5. Check blue and red tubing for kinks and occlusions. NOTE: lf a kink or occlusion is found in the blue tubing, it is recommended to QC the RBC product to ensure no hemolysis occurred. 6. Touch Continue to continue

Solution

Solution steps are available to members and are not shown on this page. Answer the questions to open the solution.

Remote technician support

Get help from a technician over a video call.