GE Healthcare mamografi cihazı All models code RS232

No approved translation for this record yet; showing the original text.

Meaning

1. The 250 Series Monitor Serial Interface is capable of being tested by means of a loop back connector. A loop back connector simply routes the TX line from the serial port to the RX line of the same serial port. 2. Enter the Install Options Screen 1. (“How do I access the service setup / install options screens?” on page 5-27.) 3. In the Install Options Screen 1, rotate the Trim Knob to highlight the COMM softkey at the bottom of the display below the menu bar. 4. Press the Trim Knob to enter the Communications Setup screen. 5. On the Communications Setup screen, the three items in the column at the far left represent the rear panel RS232 Communications Port designations J109, J110 and J111. The middle column represents the current Baudrate for the associated communications port (same line reading across), and the last column on the right represents the communication protocol for the associated communication port. 6. To change a Baudrate, rotate the Trim Knob to highlight the desired baudrate selection on the desired communication port line. 7. Press the Trim Knob. The current setting displays in blinking inverse video. 8. Now rotate the Trim Knob to change the current baudrate s

Member details