Error code
Baxter All models code IF01
No approved translation for this record yet; showing the original text.
Meaning
Pump software not compatible with hardware revision
Member details
Signed-in members see the LED pattern and how to read the code with the remote control on eligible records.
View details as a memberPossible causes
Replace software
Solution
Solution steps are available to members and are not shown on this page. Answer the questions to open the solution.
Remote technician support
Get help from a technician over a video call.
Address of this page: https://www.remfixit.com/tr/baxter/tum-modeller/if01